1、 泄露電流測試 / Leakage Current Measurement
測試檔位:
*注意:泄露電流測試在設備通電運行下測試
(1) 對地漏電流IPE / EARTH LEAKAGE CURRENT:
電流從電源通過絕緣體到保護導體,然后流到大地。/Current which flows from the power pack over the insulation to the protective conductor, and thus to earth.
測試方法:將被測設備電源插頭插到測試儀SECULIFE ST側面插座,按回車鍵/確認鍵開始測試,即可顯示測試結果。
(2) 接觸電流/外殼泄露電流 IHL / Contact Current:
電流從沒有連接保護導體的外殼通過外部導體連接到大地或別的外殼部件。/Current which flows from housing parts which are not connected to the protective conductor via an external conductive connection to earth or another part of the
housing. Flow of current via the protective conductor is excluded in this case.
The AC component is measured.The DC component can also be measured if individual measurement is performed (instead of a test sequence).
測試方法:將被測設備電源插頭插到測試儀SECULIFE ST側面插座,按回車鍵/確認鍵開始測試,測試表筆接觸金屬外殼外漏可能導電部分,即可顯示測試結果。
2、 保護導體電阻測試/接地連接一致性 Measuring Protective Conductor Resistance
測試檔位:
在被測裝置的安全接地系統中,必須要保證電氣產品的接地部分通過接地導體牢靠地連接到接地點。同樣也要確認電氣裝置的底座和其它容易接觸到的導電部分牢固地連接到安全接地點。IEC60601標準規定的測試電流是25A,IEC62353規定的測試電流是>200mA。
測試方法:將CLASS I 類設備電源插頭插到測試儀SECULIFE ST側面插座,測試表筆一旦接觸到設備各個接地點,便自動按200mA測試電流開始測試,并顯示結果;如需要切換到25A測試電流,按向上▲鍵即可切換。

Maximum Allowable Limit Values for Protective Conductor Resistance for Connector Cables with a Length of up to 5 m:
3、 絕緣測試/耐壓測試 Insulation Measurement/High-Voltage Test:
測試檔位:
(1) 絕緣測試 / Insulation Measurement:
把測試電壓加在正常情況下帶電的導體部分(L和N內部短接)與PE或者其它可接觸到的導電部分之間(II類儀器)不能有出現擊穿或者連續的電火花現象。
測試方法:將被測設備電源插頭插到測試儀SECULIFE ST側面插座,按回車鍵/確認鍵開始測試,即可顯示測試結果。
(2) 耐壓測試 High-Voltage Test:
CLASS I類設備:測試儀將被測設備L和N內部短接,在與PE之間施加高壓測試。
CLASS II類設備:測試儀將被測設備L和N內部短接,在與測試表筆接觸位置之間施加高壓測試。
測試方法:將CLASS I類設備電源插頭插到測試儀SECULIFE ST側面插座,設置等效的交流測試電壓(實際輸出1.5倍的直流電壓),一直按住回車鍵/確認鍵開始測試,并60s倒計時。在所需測試時間內,正常實施設定的測試電壓,未自動結束測試,或者未發現電火花,則表示測試結果OK。
1、 泄露電流測試 / Leakage Current Measurement
測試檔位:
*注意:泄露電流測試在設備通電運行下測試電流從電源通過絕緣體到保護導體,然后流到大地。/Current which flows from the power pack over the insulation to the protective conductor, and thus to earth.
測試方法:將被測設備電源插頭插到測試儀SECULIFE ST側面插座,按回車鍵/確認鍵開始測試,即可顯示測試結果。

(2) 接觸電流/外殼泄露電流 IHL / Contact Current:
電流從沒有連接保護導體的外殼通過外部導體連接到大地或別的外殼部件。/Current which flows from housing parts which are not connected to the protective conductor via an external conductive connection to earth or another part of the
housing. Flow of current via the protective conductor is excluded in this case.
The AC component is measured.The DC component can also be measured if individual measurement is performed (instead of a test sequence).
測試方法:將被測設備電源插頭插到測試儀SECULIFE ST側面插座,按回車鍵/確認鍵開始測試,測試表筆接觸金屬外殼外漏可能導電部分,即可顯示測試結果。

測試檔位:

測試方法:將CLASS I 類設備電源插頭插到測試儀SECULIFE ST側面插座,測試表筆一旦接觸到設備各個接地點,便自動按200mA測試電流開始測試,并顯示結果;如需要切換到25A測試電流,按向上▲鍵即可切換。



3、 絕緣測試/耐壓測試 Insulation Measurement/High-Voltage Test:
測試檔位:

(1) 絕緣測試 / Insulation Measurement:
把測試電壓加在正常情況下帶電的導體部分(L和N內部短接)與PE或者其它可接觸到的導電部分之間(II類儀器)不能有出現擊穿或者連續的電火花現象。
測試方法:將被測設備電源插頭插到測試儀SECULIFE ST側面插座,按回車鍵/確認鍵開始測試,即可顯示測試結果。

(2) 耐壓測試 High-Voltage Test:
CLASS I類設備:測試儀將被測設備L和N內部短接,在與PE之間施加高壓測試。
CLASS II類設備:測試儀將被測設備L和N內部短接,在與測試表筆接觸位置之間施加高壓測試。
測試方法:將CLASS I類設備電源插頭插到測試儀SECULIFE ST側面插座,設置等效的交流測試電壓(實際輸出1.5倍的直流電壓),一直按住回車鍵/確認鍵開始測試,并60s倒計時。在所需測試時間內,正常實施設定的測試電壓,未自動結束測試,或者未發現電火花,則表示測試結果OK。

件號M6930的多功能測試儀器,SECULIFE ST符合EN 62353,EN 60601,EN 61010,EN 60335,EN 60950,GOST R(ГОСТР)60601,VDE 0701-0702的測試程序的測試儀,非常適合在醫療設備上執行和記錄測量。它可以對便攜式醫療設備進行精確測試,也可以用于大型設備,例如X射線設備和MRI掃描儀。